Overview
Bonvera Express 5R is an advanced 5-Part Differential Hematology Analyzer designed for CBC + 5 PDA, CBC + 5 PDA + Retic, and Body Fluid testing. It offers 28 reportable and 17 research parameters, high throughput of 90 samples/hour in CBC mode, 4-angle laser scattering technology for precision and economy, and a micro sample requirement of just 20 µL. With only 4 reagents plus 1 dye tube for Retic tests, 500,000-result storage, and eco-friendly sheath liquid usage, it’s ideal for reliable, efficient, and high-end hematology testing.
Specifications
WBC 5-part flow cytometry (FCM)+4 angles laser light scatter
WBC analysis: optical and impedance measurements
RBC/PLT analysis: impedance method
HGB test: cyanide-free reagent colorimetry
RETIC test: flow cytometry (FCM) + laser light scatter + chemical dye method
Parameters
WBC, LYM%, MON%, NEU%, EOS%, BAS%, LYM#, MON#, NEU, EOS#, BAS#, RBC, HGB, HCT, MCV.
MCH, MCHC, RDW_CV, RDW_SD, PLT, MPV, PDW, PCT, P_LCR, PLCC, RETIC%, RETIC ABS, IRF
7 body-luid parameters: WBC-BF, RBC-BF, MN%, MN#, PMN%, PMN#, TC-BF#
14 research parameters: ALY%, ALY#, LIC%, LIC#, *NRBC%, *NRBC#, *BLAST%, *BLAST#, InR%, InR#, Eos-BF%, Eos-BF#, Neu-BF%, Neu-BF# (*under development & study)
Blood routine provide: 3 histograms for WBC, RBC and PLT
2 scatter grams: 5-part differential scatter gram, eosinophils and neutrophils scatter gram
Two 3D stereograms
RETIC provides: 2 histograms, 2 scatter grams, two 30 stereograms
Body fluid provides: 2 histograms, 2 scatter grams, two 3D stereograms
Sample Modes
Auto loader 20µL, manual mode 20µl, pre-diluted mode 20µl, body-fluid mode 70µl
Throughput
CBC mode: 90T/H, CBC+5DIFF mode: 80T/H, Body-fluid mode: 30T/H
Aperture Diameter
WBC 100µm, RBC/PLT 68µm
Sampling Device
Auto loader with 50 (10 x 5 racks) sample storage capacity | Built-in bar code reader (Optional)
Data Storage
Up to 500,000 sample results with all data & graphics
FAQ
## What is Bonavera Express 5R?
Bonavera Express 5R is a new generation high-end Hematology analyzer providing 49 CBC parameters including 28 reportable CBC parameters with 5-part WBC differential, Reticulocyte enumeration, 7 Body Fluid parameters, and 17 RUO parameters.
## Which Reticulocyte Parameter results Bonavera Express 5R Provides?
It provides results for Retic_Abs count, Retic% and IRF.
## What is Body Fluid Analysis?
Body fluid analysis helps diagnose various conditions by examining cell counts and characteristics in fluids like synovial, pleural, peritoneal, and pericardial fluids, aiding in detection of inflammation, infection, trauma, and malignancy.
## Bonavera Express 5R provides results for which Body Fluid Parameters?
It provides 7 Body Fluid parameters: Total Count-BF, WBC Count-BF, RBC Count-BF, Mononuclear Cells-BF (No & %), and Polymorphonuclear Cells-BF (No & %).
## Which are the RUO Parameter results offered by Bonavera Express 5R?
For Blood: Atypical Lymphocyte (ALY) # & %, Large Immature Cells (LIC) # & %, Nucleated RBC (NRBC) # & %, Blast Cells # & %, and Infected RBC (InR) # & %. For BF samples: Eosinophil-BF # & % and Neutrophil-BF # & %.
## Can Bonavera Express 5R indicate Malaria Parasite?
Yes, parameters InR# and InR‰ help detect malaria-infected red blood cells by indicating count and proportion of infected cells.
## What is the speed of testing offered by Bonavera Express 5R?
It offers 90 samples/hour in CBC mode and 80 samples/hour in CBC + 5 PDA mode.
## Which sample operational modes are available?
Whole Blood Mode, Capillary Mode, and Prediluted Mode.
## How much minimum sample volume is required?
Minimum 20µl of sample is required.
## Does Bonavera Express 5R provide graphical counting results?
Yes, it provides histograms, scattergrams, and stereograms for CBC, Reticulocyte, and Body Fluid analysis.
## How many results can be stored?
It can store up to 5,00,000 results including graphs and scattergrams.
## How many reagents are used?
It uses 4 routine reagents (Diluent, Lyse, Detergent, Sheath) and 1 maintenance reagent (Probe Cleaner). Reticulocyte counting requires an additional Retic prefilled tube.
## What is Sheath reagent?
Sheath reagent preserves WBC morphology for accurate optical scatter measurement and is dye-free, making it eco-friendly.
## What distinct advantage does it offer over other analyzers?
It uses Flow Cytometry with 4-angle laser scatter separation, offering more accurate WBC differentiation at economical cost.
## What about the display of Bonavera Express 5R?
It comes with a separate 17” color display allowing easy visualization of results, graphics, and flags on a single screen.
## What sampling devices are offered?
It supports single-sample operation with cap piercing and also includes an Autoloader for multiple samples.
## What is the capacity of the Autoloader?
The autoloader can process 50 samples at a time using 5 racks of 10 tubes each.
## Can it be connected to LIS/HIS?
Yes, it supports LIS/HIS connectivity and follows HL7 protocol.